CAS
750 Multiparameter Monitor
Designed with the leading technology
you trust, this compact, portable monitor is ideal
for a variety
of patient monitoring applications including emergency
care, transport, in-hospital applications, outpatient
care and procedural sedation.
Small, Lightweight and Technologically Advanced.
At 4.4 lbs (2kg), the CAS 750 is one of the world’s
smallest and lightest multi-parameter monitors. It
was especially designed for the way you work today.
With this one small unit, you can monitor all the
key parameters quickly and easily. Moving from one
location to another is never a problem. CAS has
incorporated only best-in-class technologies so you
can count on unmatched reliability, accuracy and
performance.
Features
• Fully
configured monitor with ECG, Respiration, Temperature,
SpO2, NIBP and EtCO2, weighing only 4.4 lbs (2
kg)
•
Pulse oximetry choices including Masimo SET® or
Nellcor® OxiMax®
•
MAXNIBP® (Motion Artifact Extraction) for fast,
accurate and reliable blood pressure measurements
•
Oridion MicroStream® EtCO2 option for intubated
and non-intubated patients
• Wireless (infrared) printer option
• Monitoring for Neonates to Adults
• User removable NiMH battery pack, internal AC power
supply and DC power option.
Two
main variants of the CAS 750 are available:
| |
 |
| |
750E
Configuration
750E-1 ECG
750E-2 ECG and SpO2
750E-3 ECG, SpO2 and MAXNIBP
750E-3C ECG, SpO2 and EtCO2
750E-4 ECG, SpO2, MAXNIBP and EtCO2 |
| |
 |
| |
750C
Configuration
750C-2 EtCO2 and SpO2
750C-3 EtCO2, SpO2 and MAXNIBP
|
(Each
of the above is also available with an external
infra-red printer).
|
SPECIFICATIONS
ECG
Input: 3 or 5 leads
Leads: I, II, III, aVL, aVF, aVR, V
Frequency Response: 0.5 to 40 Hz
Gain: Auto or Selectable (X1, X2,
X4 or X8)
Pacer Detection: Yes
Pacer Rejection: Yes
Heart Rate (ECG)
Range: Adult 15-300 BPM, Neonate 15-350
BPM
Accuracy: ± 3 BPM or ± 3%,
whichever is greater
Resolution: 1 BPM
Respiration (ECG)
Method: Impedance Pneumography
Range: 0 – 150 BrPM
Accuracy: 0 to122 BrPM ± 1 BrPM,
123 to 150 BrPM ± 2 BrPM
Resolution: 1 BrPM
Temperature
Input: YSI® 400 compatible
Range: 29 °C to 42 °C
(84.2 °F to 107.6 °F)
Accuracy: ± 0.1 °C (± 0.2 °F)
Resolution: 0.1 °C or 0.1 °F
NIBP
Technique: Oscillometric (MAXNIBP® Technology)
Patient Range: Adult – Neonate
Systolic: Adult 30-255 mmHg, Neonate
30-135 mmHg
Diastolic: Adult 15-220 mmHg, Neonate
15-110 mmHg
MAP: Adult 20-235 mmHg, Neonate 20-125
mmHg
Pulse Rate Range: Adult 30-240 BPM,
Neonate 40-240 BPM
Accuracy: ± 5 mmHg. Standard
deviation no greater than 8 mmHg (meets AAMI 2002 SP-10
requirements)
SpO2
Masimo SET®
Range: 0-100%
Accuracy: Adult 70-100%, ± 2
digits (1 S.D.), Neonate 70-100%, ± 3 digits
(1 S.D.)
Pulse Rate Range: 25-240 BPM
Accuracy: ± 3 BPM
Nellcor® OxiMax®
Range: 1-100%
Accuracy: Adult 70-100%, ± 2
digits (1 S.D.), Neonate 70-100%, ± 3 digits (1
S.D.)
Pulse Rate Range: 20-240 BPM
Accuracy: ± 3 digits
EtCO2
Method: Sidestream Capnography (Oridion
MicroStream®)
Units: mmHg, % or kPa
Sample Rate: 50 ml/min ± 7.5
ml/min
Range: 0 to 99 mmHg (0-13.2 kPa and
0-13.0 vol% at sea level)
Warm-Up Time: 30 seconds typical, reaches
steady state accuracy 20 minutes after power up
Accuracy: 0-38 mmHg ± 2 mmHg,
39-99 mmHg ± 5% of reading
Resolution: 1 mmHg, 0.1 % or 0.1 kPa
Respiration Rate: 0-150 BrPM
Patient Alarms
Adjustable Alarms (High & Low): Heart
Rate, %SpO2, Pulse Rate, EtCO2, Respiration Rate, Systolic,
Diastolic, No Respiration
Fixed Alarms: Asystole, FiCO2 (> 3
mmHg)
Alarm History: 25 most recent alarms
Indicators: Audible, Yellow Equipment
Alarm LED’s, Red Patient Alarm LED’s, and
Message Window
Operating Modes
Patient Modes: Neonate or Adult
NIBP Modes: Manual, STAT or automatic (at preset intervals)
Other Modes: Stand-by mode and Sleep
Mode
Display
Numerics: HR, PR, %SpO2, NIBP, (Systolic,Diastolic
and MAP or PR), EtCO2 (mmHg, % or kPa), RR, Temp (in °F
or °C)
Waveforms: Configurable (up to 3) ECG,
SpO2, Impedance Resp, CO2
Sweep Speed:
12.5, 25, 50 mm/sec (ECG, SpO2)
3.0, 6.25, 12.5 mm/sec (Resp, CO2)
Physical Dimensions
Fully configured unit with battery
H x W x D: 6.75 in x 8.5 in x 4.0 in
(17.0 cm x 21.5 cm x 10.2 cm)
Weight: 4.4 lbs (2.0 kgs)
Operating Environment
Operating Temperature: 32 °F to
122 °F
(0 °C to 50 °C)
Storage Temperature: -4 °F to 140 °F
(-20 °C to 60 °C)
Humidity: 15 to 95% RH, non-condensing
Barometric
Pressure: 600 to 790
mmHg (80 to 105 kPa)
Power
Source: External AC power or internal
battery
AC Power: 100-240 VAC, 50/60 Hz, 0.5A;
0.5A fuse
DC (Optional): 12 VDC
Battery: Smart NiMH battery pack (user
removable)
Charge Time: 3-5 hrs
Operating Time: 3 hrs (750E-4 with NIBP
at 15 minute intervals)
Chassis Leakage Current: 100 µA
(maximum)
Standards
ANSI/AAMI SP-10: 2002
EN 60601-1, EN 60601-1-2, EN 60601-2-27, EN 60601-2-30,
EN 60601-20-49, EN 864, EN 865
UL Listed – UL 60601-1, CAN/CSA C22.2 No. 601.1
CE Marking according to Directive 93/42/EEC
|