Mindray: BeneView Request DemonstrationDownload DatasheetSend for more information

blob BeneView T6/T8: no-compromise monitors...

• 15"/17"colour TFT display
• Suitable for adult, paediatric and neonatal patients
• Up to 3 slave screens on the main units for separate view
• 13 module slots for flexible configuration
• MPM module with up to 8-hour data saving for patient transport
• Advanced technologies including RM, ICG, BIS, etc.
• Maximum 8 channels for IBP measurement
• Friendly and configurable user interface
• Powerful data review:
• Maximum 120-hour graphic and tabular trends of all parameters
• 24-hour full disclosure waveforms review
• 120 seconds frozen waveforms
• Up to 1000 pieces of NIBP record storage
• 100 pieces of regular alarms review
• 100 pieces of ARR alarms storage
• Ethernet or Wireless LAN networking

  Why Pay More?

blob Standard Configuration:

• Main unit + Display + MPM (ECG, SpO2, Dual-IBP, 2-TEMP, NIBP)


blob Optional:

All-in-one Module: MPM with optional configurations

Single Modules: 12 lead ECG module, Dual-IBP module, CO module, EtCO2 module, Multi-gas/O2 module(with optional BIS), ICG module, BIS module, Spirometry module, Thermal recorder module

Module Racks: 8 slot module rack

Navigating options: touch screen, mouse, keyboard, or Bluetooth presenter.

blob Other options:

Wireless LAN, Power-off Storage, Laser Printer, External Displays

  BeneView

Technical Specifications

Safety
Meets the requirements of IEC60601 series,

Dimension and Weight

Dimension: 400mm(W)x370mm(H)X193mm(D)
Weight:10.5 (standard configuration)

Operational Environment

Power requirements: AC100-240(±10%) 50/60Hz(±3Hz),2.5-1.4A Temperarure: 5-40oC Humidity: 15-95%

Patient Range
Adult, pediatric and neonatal

Performance Specifications

Display: 15"/17"(T6/T8)color XGA/SXGA
Support up 4 independent displays
Multi slave display selectable
Trace: Waveform channel per display 10/12(T6/T8)
More than 40 waveforms with
Independent displays
Sweep speed: 6,25mm/s, 12.5mm/s, 25mm/s, and 50mm/s
Indicator: Dual alarm indicator lights
Power indicator light
Battery idicator light
QRS beep and alarm sound
Operating key sound
Interface: Parameter module frame and interface
AC Power input socket
Network interface
External display interface (DVI)
Aux output interface
Defib sync interface
Peripheral USB Device
(mouse, keyboard, Bluetooth presenter, etc)
Battery backup: Rechargeable battery Li-ion battery
2hours working time
Trend time: 120 hours trending
Up to 24 hours full disclousure
Alarm: User-adjustable High and low limits
Prioritized audiable and visual alarm
Networking: Connected to central monitoring system
by hardwire/wireless
Bedside device connection
Connected to Clinical Information System
(Currently only available in China)
Recorder: Built-in, thermal array
Waveform: 3 channels
Printer: External printer supported

ECG

5-lead and 3 lead selectable, 12-lead optional
Input: 10-leadwire: RA;LA;RL;LL;V1-V6
5-leadwire cable: RA;LA;RL;LL;V
3-leadwire cable: RA;LA;LL
Lead selection: 12 lead: I,II,III;aVR,aVL,aVF,V1-V6
5-lead: I,II,III;aVR,aVL,aVF,V(n)
3-lead: I,II,III
ECG waveform: Up to 8 channels, 12 lead display
Gain selection: x0.125;x0.25;x0,5;x1;x2;auto
Sweep speed: 12,5mm/s, 25mm/s, 50 mm/s
Heart Rate range: Adult: 15-300 bpm;
Neonante/Pediatric 15-350bpm
Accuracy:±1bpm or ±1%, whichever is greater
Resolution: 1bpm
Filter: Diagnostic mode: 0.05-150Hz
Monitoring mode: 0.5-40 Hz
Surgical mode: 1-20 Hz
Protection: Withstand 4000VAC/50Hz voltage in isolation
Against electrosurgical interference and defibrillation
Scalling signal: 1mV ±5%
Alarm range: 15-350bpm
S-T segment detection: Measurement range: -2.0mV~2.0V
Alarm range: -2.0mV~2.0mV
Arythmia analysis: yes
Alarm: Yes, audiable and visual alarm
alarm events recallable
12 lead ECG Analysis: yes

Resipiration

Method: Thoracic impedance
Operation modes: Auto/Manual
Measurement range: Adult: 0~120 rpm
Neonate/Pediatric: 0~150 rpm
Accuracy: 7-150 rpm ±2rpm or ±2% whichever is greater
0~6 rpm unspecified
Alarm: Yes, audieble and visual alarm
alarm events recallable

NIBP

Method: Automatic oscillometric
Operation modes: Manual/Automatic/Continuous
Auto measure time: Adjustable
Measurement unit: mmHg/kPa selectable
Measurement types: Systolic, Diastolic, Mean
Measurement range:
Range of systolic pressure AdultMode 40-270mmHg
Pediatric Mode 40-200mmHg
Neonatal Mode 40-135 mmHg
Range of diastolic pressure Adult Mode 10-210mmHg
Pediatric Mode 10-150mmHg
Neonatal Mode: 10-100mmHg
Range of mean pressure Adult Mode 20-230mmHg
Pediatric Mode 20-165mmHg
Neonatal Mode 20-110mmHg
Accuracy of blood pressure measurement
The mean error shall be less than ±5mmHg
The standard deviation shall be less than 8mmHg
Over pressure protection: double safety protection
Resolution: 1mmHg
Alarm: Systolic, Diastolic, Mean

SpO2
Mindray SpO2

Measurement range: 0~100%
Resolution: 1%
Accuracy: ±2%(70-100%, Adult/PediatricNon-motion)
±3%(70-100%, Neonate, non-motion)
±3%(70-100%, Adult/Pediatric/Neonate, motion)
0-69% unspecified
Alarm range: 0-100%
Pulse rate: Range: 20~254bpm
Resolution: 1bpm
Accuracy: ±3bpm(non-motion)
±5bpm (motion)
Alarm range: 20~254bpm

MASIMO SpO2

Measurement range 1-100%
Resolution: 1%
Accuracy: ±2%(70-100%, Adult/PediatricNon-motion)
±3%(70-100%, Neonate, non-motion)
±3%(70-100%, Adult/Pediatric/Neonate, motion)
0-69% unspecified
Alarm range: 1-100%
Pulse rate: Range: 25~240bpm
Resolution: 1bpm
Accuracy: ±3bpm(non-motion)
±5bpm (motion)
Alarm range: 25~240bpm

NELLCOR SpO2

Measurement range: 0~100%
Resolution: 1%
Accuracy: 2%(70-100%, MAX-A,MAX-AL,MAX-N,
MAX-P,MAX-I and MAX FAST sensors)
2,5%(70-100%, OxiCliq A, OxiCliq N, OxiCliq I sensors)
3%(70-100%,D-YS, DS-100A, OXI-A/N and OXI-P/I Sensors)
3.5%(70-100%, MAX-R, D-YSE and D-YSPD sensors)
0-69% unspecified
Alarm range: 0-100%
Pulse rate: Range: 20~300bpm
Resolution: 1bpm
Accuracy: 20-250bpm ±3bpm
251-300bpm unpecified
Alarm range: 20~250bpm

Temperature

Measurement range 0~50oC
Resolution 0.1oC
Accuracy: ±0.1oC

Invasive blood pressure (IBP)

Measurement range: -50~300mmHg

 

Channel: up to 2 channels
Pressure transducer: Sensitivity: 5mV/V/mmHg
Impedance range: 300-3000W
Pressure names: ART, PA, CVP, RAP, LAP, ICP, CPP
Resolution: 1mmHg
Accuracy: ±2% or ±1mmHg, whichever is greater(exclusive of transducer)
Alarm range: -50~300mmHg

Cardiac Output

Method: Thermodilution
Measurements range: CO;01-20L/min.
TB:23-43oC
TI:0-27oC
Resolution: CO:0.1L/min
TI: 0.1oC
TB: 0.1oC
Accuracy: CO:±5%
TB:±0.1oC
TI: ±0.1oC
Parameter output: Cardiac output
Hemodynamics calculation

EtCO2

Method: Infrared absorption
Displayed numerics: EtCO2, FICO2, AwRR
Waveforms: Capnography

Microstream CO2

CO2 range 0-99mmHg
Accuracy: 0-38 mmHg ±2mmHg
39-99mmHg ±5% of reading + 0.08% for every 1mmHg
(above 38mmHg)
Resolution: Waveform:0.1 mmHg Value: 1mmHg
Sampling rate 50ml/min-7.5+15ml/min
Initialization time 30 seconds (typical), reaches ±5%
steady-state accuracy within 3 minutes
Response time Typical value: 2.9s, including the rising
time and the delay time
(adopting the FilterLine of standard length)
Rising time: <190ms(rising from 10% to 90%)
Delay time: 2.7s (typical value)
Respiration rate 0-150 rpm
Respiration rate accuracy 0-70rpm ±1rpm
71-120rpm ±2rpm
121-150rpm ±3rpm
Modde: Adult, noenate

Sidestream CO2

CO2 measurement range 0-99mmHg
Accuracy ±2mmHg(0-40mmhg)
±5% of reading (41-76mmHg)
±10% of reading (77-99mmHg)
Sampling rate: 100 or 150 ml/min
Sampling rate accuracy 15%
Start up line <1min, once the modul starts up
it reaches ISO accuracy Mode 10 minutes after
start up, the module reaches full accuracy mode
Respiration rate 0-120rpm
Respiration rate accuracy ±2rpm(0-70rpm)
±5rpm(70rpm)
Response time <240msec(10% to 90%)
Delay time <2s (Sampling line length: 7 inches; internal
diameter:0.055 inches
sampling gas flow rate:150ml/min)

Mainstream CO2

Method: Infrared Absorpion
Measuring Mode: Mainstream
Measuremnt Range: CO2:0-150mmHg
AwRR: 0-150 rpm
Resolution:CO2:0-69mmHg; 0.1mmHg AwRR:1rpm
Accuracy: CO2 concentration:
±2mmHg (0-40mmHg)
±5% of reading (41-70mmHg)
±8% of reading (71-100mmHg)
±10% of reading (101-150mmHg)
Alarm range: Same as Measurement range

Multi-Gas/O2

Method:Infrared Absorption/Paramagnetic
Gas sorts: CO2, N2O, Des, Iso, Enf, Sev, Hal
O2 (optional paramagnetic sensor)
Measurement range: CO2:0-30%
N2O:0-105%
O2:0-105%
Enf, ISO, Hal: 0-30%
Sev:0-30%
Des:0-30%
Measurement accuracy: CO2:0-1%±0,1%, 1-5%±0.2%, 5-7%
0.3%, 7-10%±0.5%,>10%unspecified
N2O:0-20%±2%, 20-100%±3%
Des:0-1%±0.15%, 1-5%±0.2%, 5-10%±0.4%, 10-15%±0.4%
10-15%±0.6%,15-18%±1%>18unspecified.
Sev:0-1%±0.15,1-5%±0.2%, 5-8%±0.4,>8unspecified.
Enf. Iso,Hal:0-1%±0.15%, 1-5%±0.2%,>5unspecified O2;0-25%±1%,25-80%±2%;80-100±3%
Data output: Fiand ET values
Respiration rate: ±1rpm(2-60rpm)
Unspecified(61-100rpm)
Alarm: User selectable alarm limits for all Measurements
Other: Up to 5 waveforms displayed
Agent mixture detection
MAC value displayed

Bispectral Index(BIS)

Method: EEG Bispectrum analysis, DSA
BIS range: 0-100
SQI range: 0-100%
Other calculated parameters: Signal Quality Index(SQI), EMG
Suppression Ratio(SR), TP
BIS smooth rate: 15s or 30s (default)
EEG specifications: EEG scales: 6.25uV/12.5uV/25uV/50uV/100uV
Sweep speeds:12.5/25/50 mm/sec
Noise (RTI):<03uVRMS 0.25-50Hz
Input range:±1mV
EEG Bandwith: 0.25-110Hz
A/D Conversion: Noise shaped Sigma-Delta
Sample Rate: 16.384 samples/seconds
Patient Leakage:<10uA

Impedance Cardiography(ICG)
Method: Thoracicrange: Bioimpedance measurement
DISQTM (Digital Impedance Signal Qualifier) Technology
Z MARC (Impedance Modulating AoRtic Compliance) algorithm
Waveforms: Thoracic Bioimpedance
Stroke Volume(SV)range: 5-250ml/b
Heart rate (HR) range: 44-185bpm
Cardiac Output(CO)range: 1,4-15l/min
Other measured parameters: Flow, Resistance, Contractility, Fluid
Status and Cardiac Work Parameters:
SI, CI, BSA,SVR/SVRI,PVR/PVRI,LCW/LCWI,LVSW/LVSWI,RCW/RCWI
RVSW/RVSWI,EF,ACI,VI,TFC,TFI,STR,PEP,LVET,VEPT.
Alarm: User selectable alarm limits for TFC, CI

Respiratory Mechanics(RM)

Method: Flow measurement based on micro
differential pressure sensing
Waveforms: Airway Pressure, Flow, Volume, Loops (P-V,V-F)
Flow: Range: Aldult, Pediatric:±(2-120)l/min
Neonate:±(0.5-30)l/min
Accuracy:25ml/s or ±10% of reading
Whichever is greater
Resolution: 0.1l/min
Pressure unit: User selectable cmH2O, kPs, or mmHg
Airway pressure: Range:-20.0 to 120.0 cmH2O
Accuracy: ±3% of reading
Resolution: 0.1 cmH2O
Minute volume: Range: Adult, Pediatric:2-60l/min
Neonate: 0.5-15l/min
Accuracy: ±5% of reading
Tidal volume: Range: Adult, Pediatric:100-1500ml
Neonate: 20-500ml
Resolution:1ml
Accuracy:±5% of reading
Respiration Rate: Range: Adult, Pediatric: 4-99rpm
Neonate:10-99rpm
Accuracy:±2rpm
I:E ratio Range: 12.0:1 - 1:12.0
Accuracy:±5% of reading
Dead space:<11ml
Other Measured parameters: Respiratory Mechanics parameters: I:E ratio
FEV 1.0,MAP,TVi/TVe,
MVi/MVe,PEEP,PEF,PIF,PIP,C,Compliance,RAW,RSBI,NIP,WOB. Alarm: User selectable alarm limits for Respiration Rate,PEEP,PIP,Mve

blob Beneview Options

Main unit: Main unit with integrated display, power supply and navigating lever
Main unit with integrated display, touchscreen, power supply, and navigating lever
All-In-One Module: Stand-of-Care MPM modules (6 kind of options)
Parameter Modules: Dual IBP module, CO module, EtCO2 module, Multi-Gas/O2 modules(with optional BIS module),
Respiratpry Mechanics Module
Module Rack: 8 slot module rack
Navigating Options: USB compatible mouse, keyboard, Bluetooth presenter
Printing Options: 3 channel thermal recorder module , printer
Mounting Options: Rolling stand, wall mount
Other Options: External Display, Wireless LAN, Power off Storage & Extensive memory card


Model Selections

  BeneView T6 BeneView T8
Integrated Display 15" color TFT 17" color TFT
External independent display support standard: 1, Option:3 standard:3
Waveform channel Standard 10(up to 40 with independent displays) 12(up to 48 with independent displays)
Touch screen operation Optional Optional
MPM Module Compatible Compatible
External module frame Compatible Compatible
IBP measurement Up to 8 channels Up to 8 channels
Data trending Up to 120 hrs Up to 120 hrs
Full disclousure review & Power-off storage Optional Standard
Calculations Standard: 2 kinds, Drug & Homodynamic Calculation Standard: 5 kinds, Drug, Homodynamic, Oxygen, Ventilation, Nephritic

MPM Module Options:

  M1 M2 M3 M4 M5 M6
3/5-lead ECG x x x x x x
12-lead ECG       x x x
RESP x x x x x x
Dual channel TEMP x x x x x x
NIBP x x x x x x
Dual channel IBP x x x x x x
Mindray SPO2 x     x    
Masimo SPO2   x     x  
Nellcor SPO2     x     x

Multi-Gas/O2 Module Option

  GAS1 GAS2 GAS3 GAS4 GAS5 GAS6 GAS7 GAS8
Auto-ID AG x x x x        
Manual-ID AG         x x x x
O2 x x     x x    
BIS x   x   x   x  

NOTE: specifications subject to change without prior notice