Technical
Specifications
Safety
IEC60601 - 1 approved
Dimension and Weight
Dimension: 261mm (W) x 240mm(H) x 171mm(D)
Weight: <5kg (unincluding battery)
Operational Environment
AC Power Requirements: 100-240 VAC(±10%), 50/60Hz(±3Hz), 110VA
DC Power Requirements 12VDC (rated), 10-16VDC 70W
Temperature: 0-40oC
Humidity: 15%-95%, non-condensing
Patient Range
Adult, pediatric and neonatal
Performance Specifications
Display: 8,4" color TFT (diagonal)
Rolling and refreshing waveform display
Resolution 800x600
Multi displays selectable, including:
Standard display
Large-font Display
Trend Coexist Display
Freeze Display
Alarm limit display
Multi - lead ECG simultaneous display
Bed-to-bed view display
OxyCRG dynamic View display
Trace: 8 waveforms
Sweep speed: 6,25mm/s, 12.5mm/s, 25mm/s, and 50mm/s
Indicator: Alarm indicatoe light
Power indicator light
Battery indicator light
QRS beep and alarm sound
Interface: Parameter cable interface
AC and DC Power input socket
Software upgrade through networking port
External VGA interface for connections
of an alternate display
Aux Output (Analog/Defib-Synch/Nurse Call)
Li-ion Battery: Rechargeable, maximum 6.5 hours for charging,
3 hours for continous working
Lead Acid Battery: Rechargeable, maximum 6 hours for charging:
75 minutes for continous working
Trend-time: 1~96 hours
Alarm: User adjustable High and Low limits 3-level audible
And visual alarm
Networking: Connected to central monitoring system
Recorder: Built-in, thermal array
Plethysmogram waveform: 2 channels
Record mode: manual, on alarm, time-defined, etc.
Paper width: 50 mm
Print speed: 25mm/s, 50mm/s
ECG
5-lead or 3-lead selectable
Input: 5-lead: RA;LA;RL;LL;V
Lead selection: I;II;III;aVR;aVL;AVF;V
ECG waveform: 2 channels
Gain selections: x0.125;x0.25;0.5;x1;x2;auto
Sweep speed: 12.5mm/s, 25mm/s,50mm/s
Heart Rate range: Adult: 15-300bpm; neonate/pediatric: 15-350bpm
Accracy: ±1bpm or ±1%, whichever is greater
Resolution: 1bpm
Filter: Diagnostic Mode: 0.05-100Hz
Monitorig mode: 0.5-40 Hz
Surgical mode: 1-20 Hz
Protection: Withstand 4000VAC/50Hz voltage in isolation
Against electrosurgical interference and
defibrillation
Scalling signal: 1mv±5%
Alarm range: 15-350bpm
S-T segment detection: Measurement range: -2.0mV~2.0mV
Alarm Range: -2.0mV~2.0mV
Arrhythmia analysis: YES
Standard Configuration
ECG+RESP+Dual-TEMP, SPO2, NIBP, Lead acid battery
Pace detection: YES
Alarm: YES, audibble and visual alarm, alarm events
recallable
Respiration
Method: Thoracic impedance
Sensing leads: Lead I and lead II are optional(default:leadII)
Operation modes: Auto/Manual
Measurement Range: Adult: 0~120 BrPM; Neonate/Child 0~150 BrPM
Apnea alarm: YES
NIBP
Method: Automatic oscillometric
Operation modes: Manual/Automatic/Continous
Auto measure time: Adjustable
Measurement unit: mmHg/kPa selectable
Measurement types: Systolic, Diastolic, Mean
Measuremnt Range:
Range of systolic pressure AdultMode 40-270mmHg
Pediatric Mode 40-200mmHg
Neonatal Mode 40-135 mmHg
Range of diastolic pressure Adult Mode 10-210mmHg
Pediatric Mode 10-150mmHg
Neonatal Mode: 10-100mmHg
Range of mean pressure Adult Mode 20-230mmHg
Pediatric Mode 20-165mmHg
Neonatal Mode 20-110mmHg
Accuracy of blood pressure measurement
The mean error shall be less than ±5mmHg
The standard deviation shall be less than 8mmHg
Over pressure protection: double safety protection
Resolution: 1mmHg
Alarm: Systolic, Diastolic, Mean
Accuracy: mean error ±5mmHg
Standard deviation 8mmHg
|
|
SpO2
Mindray SpO2
Measurement range: 0~100%
Resolution: 1%
Accuracy: ±2%(70-100%, Adult/PediatricNon-motion)
±3%(70-100%, Neonate, non-motion)
±3%(70-100%, Adult/Pediatric/Neonate, motion)
0-69% unspecified
Alarm range: 0-100%
Pulse rate: Range: 20~254bpm
Resolution: 1bpm
Accuracy: ±3bpm(non-motion)
±5bpm (motion)
Alarm range: 20~254bpm
MASIMO SpO2
Measurement range 1-100%
Resolution: 1%
Accuracy: ±2%(70-100%, Adult/PediatricNon-motion)
±3%(70-100%, Neonate, non-motion)
±3%(70-100%, Adult/Pediatric/Neonate, motion)
0-69% unspecified
Alarm range: 1-100%
Pulse rate: Range: 25~240bpm
Resolution: 1bpm
Accuracy: ±3bpm(non-motion)
±5bpm (motion)
Alarm range: 25~240bpm
NELLCOR SpO2
Measurement range: 0~100%
Resolution: 1%
Accuracy: 2%(70-100%, MAX-A,MAX-AL,MAX-N,
MAX-P,MAX-I and MAX FAST sensors)
2,5%(70-100%, OxiCliq A, OxiCliq N, OxiCliq I sensors)
3%(70-100%,D-YS, DS-100A, OXI-A/N and OXI-P/I Sensors)
3.5%(70-100%, MAX-R, D-YSE and D-YSPD sensors)
0-69% unspecified
Alarm range: 0-100%
Pulse rate: Range: 20~300bpm
Resolution: 1bpm
Accuracy: 20-250bpm ±3bpm
251-300bpm unpecified
Alarm range: 20~300bpm
Temperature
Measurement range 0~50oC
Resolution 0.1oC
Accuracy: without sensor
Invasive blood pressure (IBP)
Measurement range: -50~300mmHg
Channel: 2 channels
Pressure transducer: Sensitivity: 5uV/V/mmHg
Pressure names: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Resolution: 1mmHg
Accuracy: ±2% or ±1mmHg, whichever is greater
(exclusive or transducer)
Alarm Range: -10~300mmHg
EtCO2
Microstream CO2
CO2 range 0-99mmHg
Accuracy: 0-38 mmHg ±2mmHg
39-99mmHg ±5% of reading + 0.08% for every 1mmHg
(above 38mmHg)
Resolution 1mmHg
Sampling rate 50ml/min-7.5+15ml/min
Initialization time 30 seconds (typical), reaches ±5%
steady-state accuracy within 3 minutes
Response time Typical value: 2.9s, including the rising
time and the delay time
(adopting the FilterLine of standard length)
Rising time: <190ms (rising from 1-% to 90%)
Delay time: 2.7s(typical value)
Respiration Rate: 0-150 breaths/min
Respiration Rate Accuracy 0-70BrPM ±1BrPM
71-120BrPM ±2BrPM
121-150BrPM ±3BrPM
Mode: Adult, Neonate, Pediatric
Sidestream CO2
CO2 measurement range 0-99mmHg
Accuracy ±2mmHg (0-40mmHg)
±5% of reading (41-76mmHg)
±10% of reading (77-99mmHg)
Sampling rate 100, 150ml/min
Sampling rate accuracy 15%
Start-up time <1min, once the module starts up,
it reaches ISO Accuracy Mode
10 miinutes after start-up, the module reaches full accuracy mode
Respiration rate 0-120BrPM
Respiration rate accuracy ±2BrPM (0-70BrPM)
±BrPM(>70BrPM)
Response Time <240msec (10% to 90%)
Delay time <2s
Sampling line lenght: 7inches; internal diameter:
0.055 inches;
Sampling gas flow rate:150ml/min |