Technical
Specifications
Safety
Meet the requirements of IEC60601 series, CE marking according to MDD93/42/EEC
Dimension and Weight
Dimension: 170mm(W)x240mm(H)170mm(D)
Weight: <3.5 kg (including battery)
Operation Environment
AC Power Requirements: 100-240VAC(±10%), 50/60Hz(±3Hz),70VA
Temperature: 0-40oC(SmarTemp: 10-40oC)
Humidity: 15%-95%, non-condensig
Patient Range
Adult, pediatric, neonatal
Performance Specifications
Display:
7 segment LEDs for systolic, diastolic, mean pressures, SPO2, Temp, and pulse rate
3.4"FSTN LCD Resolution: 320x160 Pitch 0.24x0.24mm
Trace: 1 plethysmogram waveform
Sweep Speed: 25m/s
Indicator: Alarm Indicator
Power Indicator
Battery Indicator
Patient Type Indicator
Pulse Strenght Indicator
NIBP Status Indicator
Silence Indicator
Interface: Network port
Nurse call interface
Trend: Flexible database lists trend for review of up to 1200
measurements on up to 100 patients
Alarm: 3-level audible and visual alarm
Networking: Connected to Mindray central monitoring system or data
exporting system
Recorder: Built-in, thermal array
Record with: 48mm(record paper 50mm)
Print speed: 25mm/s
Lead Acid Battery: Rechargeable
About 8 hours (Charging time)
About 4.5 hours (working time)
Li-ion Battery: Rechargeable
About 8 hours (Charging time)
About 10.5 hours (Working time)
NIBP
Method: Automatic oscillometric
Operation modes: Manual/Automatic/Continual
Auto measure time: Adjustable
Measurement unit: mmHg/kPa selectable
Measurement types: Systolic, Diastolic, Mean, Pulse Rate
Measurement range: Adult:10-270 mmHg
Pediatric: 10-200mmHg
Neonate: 10-135mmHg
Over-pressure protection: Yes
Resolution: 1 mmHg
Alarm: Systolic, Diastolic, Mean, Pulse Rate
Pulse Rate Range: 40-240 bpm
Pulse Rate Resolution: 1bpm
Pulse Rate Accuracy: ±2bpm
|
|
Mindray SPO2
Measurement range: 0-100%
Resolution: 1%
Accuracy: Adult/Pediatric: ±%2%(70-100%);
Neonatal:±3(70-100%);
0-69%unspecified
Alarm range: 0-100%
Pulse rate: Range: 20-254 bpm
Resolution: 1bpm
Accuracy:±3bpm
Alarm range: 20-254bpm
Masimo SET SpO2
Measurement range: 1-100%
SpO2 accuracy:
No motion conditions Adult/Pediatric:70%-100%±2%;0-69%unspecified
Neonate: 70%-100%±3%;0-69% unspecified
During motion conditions 70%-100%±3%;0-69%unspecified
Low perfusion performance:
Pulse amplitude:>0.02%
Light transmissivity:>5%
SpO2 accuracy:±2%
PR accuracy: ±3bpm
Pulse Rate: Range 25-240bpm
Accuuracy In static condition:±3bpm
In motion condition:±5bpm
Nellcor Oximax SpO2
Measurement range: 0-100%
Compatibile sensors type: OxiMax series MAX series MAX-A,
MAX-AL, MAX-N, MAX-P, MAX-I, MAX-FAST,
MAX-R, OxicliqA, OxiCliqN, OxiCliqP, OxiCliqI,
D-YS, D-YSE, D-YSPD, DS100A, OXI-A/N, and OXI-P/I
SpO2 accuracy: 70%-100%±2%;0-69% unspecified
(sensors: MAX-A, MAX-AL, MAX-N, MAX-P, and MAX-FAST)
70%-100%±2.5%;0.69%unspecified
(sensor: OxiCliqA, OxiCliqN, OxiCliqP and OxiCliq I)
70%-100%±3%,0-69 unspecified
(sensors:DY-S,DS-100A,OXIA/N and OXI-P/I)
70%-100%±3.5%;0-69% unspecified
(sensors: MAX-R, D-YSE and YSPD)
Pulse Rate:
Range and accuracy 20-250±3bpm
251-300 bpm unspecified
SmarTemp Thermometer
Measurement range 25-44oC (77-111.2oF)
Resolution In MONITOR mode: 0.1oC (o.2oF)
Accuracy: In MONITOR mode: 25-32oC(77-89.6oF:±0.2oC(±0.3oF)
including 32oC (89.6oC)
In MONITOR mode: 32~44oC (89.6-111.2oF):±0.1oC(±0.2oF)
excluding 32oC(89.6oF)
Typical Measurement time <12s
|