Special Modes of Operation:
Preset Parameters
by Patient Weight
Adaptive Flow
Adaptive Time
Backup Apnea Ventilation
Sigh Breath
Adjustable Rise Time
Easy Exhale
100% O2 Suction
Control Principle: Closed
Loop Flow & Pressure Controller
Respiratory Rate 1 to
50 BPM
Tidal Volume 100 to 2,000 ml
Inspiratory Pressure Limit 5 to 80 cm H2O
Inspiratory Time Adaptive Time or 0.3 to 3 seconds
Peak Flow Adaptive Flow or 1 to 120 L/min Spontaneous
up to 180 L/min
Oxygen Mix (FiO2)* 21% to 100%
PEEP 0 to 20 cm H2O
Trigger Sensitivity Flow Sensitivity: 1 to 9 L/min + off
Pressure Sensitivity: -0.5 to -8 cm H2O + off
Pressure Support Ventilation (PSV) 1 to 60 cm H2O
Monitoring and Displayed Parameters:
Pressure and Flow Waveforms
Pressure, Flow and Volume Loops*
Airway Pressure (Analog Bar Graph & Numerical PIP)
Total Breath Rate (real time & trend)
Exhaled Tidal Volume (real time & trend)
Exhaled Minute Volume
Peak Flow
Inspiratory Time
Lung mechanics (Compliance, Resistance)*
Electrical Power Source (External / Internal)
Battery Level
* indicates optional feature not
available in all iVent 201 models
Size and Weight:
Dimensions
Height: 12.0 in / 350 mm
Width: 9.5 in / 240 mm
Depth: 11.5 in / 290mmWeight (no battery): 17.5 lb / 8 kg
Power:
Battery Weight: 6.5 lb / 3 kg
Overall Weight: 24.0 lb / 11 kgPower Supply:
External AC: 100 to 240V, 50-60Hz, Max 1.6A
External DC: 12 to 15V (Optional: 10 to 30V), Max 8.5A
Internal Battery: Sealed Lead-Acid 12V, 7.2Ah (rechargeable)
1 to 2 hours operating time (dependent on ventilation parameters)
Oxygen (Enrichment)
Supply:
High Pressure Supply: 40 to 60 psi (2.8 to 4.2 bar)
Low Pressure Supply: Max 80 L/min or 0.5 psi
User
Adjustable Alarms:
Rate (High/Low)
Minute Volume (High/Low)
Pressure (High/Low)
Apnea (5 to 120 seconds)
FiO2 (21 to 100%)
Leak (0 to 100%)
|
Additional
Alarms and Indicators:
High
I:E ratio
Low
O2 Pressure
AC disconnect
Internal
Battery use
Over
Temperature
Ventilator
Malfunction
Patient
Disconnect
Sensor
Problem
Date
and Time
External Interface:
Remote Monitor
(VGA)
RS-232 Serial
Port, 9 Pin Connector
DIN
Keyboard Connector
RJ-45
Remote Alarm Connector
Environmental
Specifications:
Operating Temperature: -15 to 50 °C / 0 to 120 °F
Storage Temperature: -15 to 70 °C / -4 to 140 °F
Relative Humidity: 15 to 95% at 30 °C / 85 °F
Water Resistance: IP54 (Splash Proof)
Atmospheric Pressure: 430 to 825 mmHg (15000 ft)
Vibration: IEC 68-2-6 and IEC 68-2-34
MIL-STD-810E
Shock: IEC 68-2-27 (100g)
MIL-STD-810E
Total External Sound Level: 40-45 dBa at one metre
Standards and Safety Requirements:
Meets the requirements of:
ISO 10651-1: Lung Ventilators for Medical use
ISO 10651-2: Particular requirements for Home Care Ventilators
ISO 10651-3: Particular requirements for Emergency and Transport
Ventilators
IEC 60601-1-1: Electrical Safety
IEC 60601-1-2: Electromagnetic Compatibility (EMC)
EN 794-1: Lung ventilators Part 1: Particular requirements
for critical care ventilators
EN 794-2: Lung ventilators - Part 2: Particular requirements
for home care use
EN 794-3: Lung Ventilators - Particular requirements for
Emergency & Transport Ventilators
ASTM F 1100-90: Standard Specification for Ventilators Intended
for Use in Critical Care
CSA C22.2 No. 601.1: Medical Electrical Equipment Requirements
for Safety
CSA C22.2 No. 601.2: Medical Electrical Equipment - Part
2: Particular Requirements for the Safety of Lung Ventilators
UL 2601.1: Medical Electrical Equipment Part 1: General Requirements
for Safety
MIL-STD-810E: Environmental Test Methods
IEC 68-2: Basic Environmental Testing Procedures
|
|